Abstract:
Speed of sound is an important quantity to characterize reference materials for ultrasonic applications, for instance. The alignment between the transducer and the test body is an key activity in order to perform reliable and consistent measurement. The aim of this work is to evaluate the influence of the alignment system to the expanded uncertainty of such measurement. A stainless steel cylinder was previously calibrated on an out of water system typically used for calibration of non-destructive blocks. Afterwards, the cylinder was calibrated underwater with two distinct alignment system: fixed and mobile. The values were statistically compared to the outof-water measurement, considered the golden standard for such application. For both alignment systems, the normalized error was less than 0.8, leading to conclude that the both measurement system (under- and out-of-water) do not diverge significantly. The gold standard uncertainty was 2.7 m∙s -1, whilst the fixed underwater system resulted in 13 m∙s -1, and the mobile alignment system achieved 6.6 m∙s -1. After the validation of the underwater system for speed of sound measurement, it will be applied to certify Encapsulated Tissue Mimicking Material as a reference material for biotechnology application. Keywords: validation, metrology, ultrasound.