dc.description.abstract |
From a commercial supplier a solution containing 226Ra has been standardized at National Laboratory for Ionizing Radiation Metrology (LNMRI) for the first time using Liquid scintillation based measurement. The measurement methods are 4[beta][alfa][omega]-[gama] live-timed anticoincidence counting. The results obtained by anticoincidence counting were adopted as reference value and its combined uncertainty was 0.94 %. The agreement of this reference value with reference system with ionization chamber was 0.56 %. This standardization was made in order to reduce the uncertainty in 226Ra measurement in Brazil. |
pt_BR |
dc.format.extent |
5 f. |
pt_BR |
dc.language.iso |
Inglês |
pt_BR |
dc.subject.classification |
Metrologia científica e aplicada |
pt_BR |
dc.subject.classification |
Incerteza de medição |
pt_BR |
dc.subject.classification |
Calibração |
pt_BR |
dc.subject.classification |
Metrologia e instrumentação em radiações ionizantes |
pt_BR |
dc.title |
226Ra activity standardization by 4[beta][alfa][omega](LS)-[gama](NaI-Tl) anticoincidence counting |
pt_BR |
dc.type |
Artigo |
pt_BR |
dc.rights.license |
É permitida a reprodução deste texto e os dados nele contidos, desde que citada a fonte. Reprodução para fins comerciais são proibidas. |
pt_BR |
bom.learningResourceType |
Artigo científico |
pt_BR |
dc.mediator |
Instituto Nacional de Metrologia Qualidade e Tecnologia (INMETRO) |
pt_BR |
dc.mediator |
Sociedade Brasileira de Metrologia (SBM) |
pt_BR |
dc.rights.type |
Termo de cessão SBM |
pt_BR |
dc.subject.decs |
Metrologia |
pt_BR |
dc.subject.decs |
Metrologia científica e aplicada |
pt_BR |
dc.subject.decs |
Medição |
pt_BR |
dc.subject.decs |
Calibração |
pt_BR |
dc.subject.decs |
Incerteza de medição |
pt_BR |
dc.subject.keyword |
Metrologia das radiações ionizantes |
pt_BR |
dc.subject.keyword |
Metrologia de radionuclídeos |
pt_BR |
dc.subject.keyword |
Método de contagem em anticoincidência |
pt_BR |
dc.subject.keyword |
Anticoincidence counting method |
pt_BR |
dc.subject.keyword |
Câmara de ionização |
pt_BR |
dc.subject.keyword |
Ionization chamber |
pt_BR |
dc.subject.keyword |
Rádio-226 |
pt_BR |
dc.subject.keyword |
Radium-226 |
pt_BR |
dc.subject.keyword |
Instituto de Radioproteção e Dosimetria (IRD) |
pt_BR |
dc.location.country |
Brasil |
pt_BR |